Determination of mercury by flameless AAS.

dc.advisorKenneth Johnsonen_US
dc.collegelasen_US
dc.contributor.authorLowrance, Gary.
dc.date.accessioned2013-01-07T22:40:00Z
dc.date.available2013-01-07T22:40:00Z
dc.date.created1974en_US
dc.date.issued2013-01-07
dc.departmentbiological sciencesen_US
dc.description54 leavesen_US
dc.description.abstractDETERMINATION OF MERCURY BY FLAME LESS AAS The procedure outlined describes an extremely sensitive and accurate method for the determination of mercury down to 50.0 ppb in solution. The procedure has been adapted to biological samples such as fish, brains and moldy wheat. The sample is taken into solution using wet digestion and an oxidizing agent. Mercury is then reduced to an elemental state and aerated from the solution in a reaction flask. The mercury vapor posses through an absorption cell of an atomic absorption spectrophotometer where its concentration is measured. The procedure is practically free from interferences except in extreme cases of organic vapors and the method shows a positive error. Gary G. Lowrance August 1974en_US
dc.identifier.urihttp://hdl.handle.net/123456789/2552
dc.language.isoen_USen_US
dc.subjectMercury.en_US
dc.subjectAtomic absorption spectroscopy.en_US
dc.titleDetermination of mercury by flameless AAS.en_US
dc.typeThesisen_US

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Lowrance 1974.pdf
Size:
2.52 MB
Format:
Adobe Portable Document Format

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
2.35 KB
Format:
Item-specific license agreed upon to submission
Description:

Collections